Product name : Dental Low Speed Handpiece 1:1 External Water Spray Contra Angle Dental Handpieces
Connection : Standard E type connection
Bur Applicable : Φ2.35mm
Cartridge : German cartridges
Color : silver
Spray : Outer water spray
Slow Speed Handpiece Quick View
1:1 External Water Contra angle handpiece
- Low Speed 22,000-27,000 rpm.
- The shank of the instrument mounted therein is 2.35 mm
- Fastening of the instruments with a turning latch
- Autoclavable 135°C
- For any lab or E-type motors
Warranty:
- Long Life-Span
- One Year Manufacturer Warranty
Packing List:
- Contra Angle handpiece: 1 each
- English manual: 1 each
1:1 External Water Handpiece Technical Data
1:1 External Water Contra Angle Handpiece Details
After each patient use , maintenance should be done as follows:
1. First remove the diamond burs , and clean up all the teeth on dental handpieces precipitate
2. Use a metal needle to clean the spray holes and fumaroles
Note: The spray holes and diamond burs into 40 -degree angle , fumaroles and diamond burs to a 30 degree angle
3. Place the clean lubricating oil is sprayed into the axial hole 1-2 seconds ( diamond burs is plugged holes )
4. In the bottom of the handpiece were ( with / without quick connector ) two kinds of styling , make the right amount of lubricant is injected into the bottom
5. The handpiece with quick connector combination ( first determine the water switch is turned off )
6. Replace the diamond burs , surrounded by white silk cloth handpiece head ( make sure the needle does not touch the diamond burs and white gauze , so as not to tie )
7. pedal depressed for about 20 seconds , look at whether there is in white gauze on any stain. If there are stains on the white gauze , again, add oil , starting from item 4 continued routine maintenance , know that white gauze until no stain
8. Finally, no more than 135 degrees Celsius temperature and high pressure sterilization
![]() |